Not All "Shockwave" Is Shockwave
By Dr. Craig Patane, D.C., FRCms · October 2026
By Dr. Craig Patane, D.C., FRCms · October 2026
"Shockwave" is printed on a lot of clinic windows now. Behind that one word sit two different technologies: true focused shockwave, which delivers a near-instant pressure spike to a point at a chosen depth, and the radial pressure wave, which drives energy in from the skin and weakens as it travels. They differ in physics, in reach and in evidence. A patient comparing clinics, or a clinician comparing machines, needs to know which one is actually in the room before anything else in the conversation means very much.
Extracorporeal magnetotransduction therapy (EMTT) is something else again. It is not acoustic at all: it is an oscillating magnetic field that reaches a broad, deep region of tissue without contact. In my clinic it is the third part of one approach. Focused shockwave, radial pressure wave and EMTT are used together, so that the whole region is prepared, the surrounding tissue is treated and the injured tissue itself is targeted. That combination is what I mean by regenerative medicine.
This article explains how they differ, what the machines on the market really produce according to their own specifications and FDA filings, where EMTT fits, and what the research does and does not show.
Shockwave is a biological treatment. It delivers mechanical energy into tissue whose healing has stalled, and the tissue responds by restarting the repair process. Too little energy and nothing happens. The right energy in the wrong place and nothing happens. The right place at the wrong depth and the skin has been treated while the tissue beneath it has not.
Before any device is switched on, the provider should be able to answer four questions about your case:
Dose matters in the trials too. In a sham-controlled trial of calcific tendinopathy of the shoulder, high-energy focused shockwave improved the Constant shoulder score by 31.0 points at six months, against 15.0 for low energy and 6.6 for sham (Gerdesmeyer et al., JAMA 2003). Total energy is necessary but not sufficient on its own: a large sham-controlled trial for plantar fasciopathy (plantar fasciitis) delivered a substantial dose and still found no difference from placebo (Buchbinder et al., JAMA 2002). Diagnosis, target and technique decide whether the energy does anything useful. I wrote more on this after the Institute for Tissue Regeneration and Repair conference: Why Shockwave Therapy Works (When It's Done Right).
This is the part most people never hear. Many clinics advertising "shockwave" are using a radial pressure wave device. Inside the handpiece, a projectile driven by compressed air or a magnet strikes a metal head resting on the skin. The pressure wave it creates is strongest at the surface and spreads and weakens as it travels inward.
A true shockwave is different physics. It reaches peak pressure in billionths of a second rather than millionths, and a focused device converges that energy on a point at a set depth inside the body, where the damaged tendon or bone actually lies. The FDA's own measurements put radial devices at 2.5 to 3.2 microseconds to peak pressure, roughly a thousand times slower than a true shockwave (more on the physics). A laboratory study of a ballistic device found its pulses too long to qualify as shock waves, with high pressure only in the first 20 to 40 mm and no focusing at all (Cleveland et al., 2007).
None of this makes radial a lesser tool. It is one of the three pillars of my multimodal approach: it produces real tissue effects that a massage gun cannot, and it treats the muscle and fascia surrounding an injury, territory that focused shockwave is not designed to cover. Without it, the protocol is incomplete. The point is only that radial and focused shockwave do different jobs, and if radial is the only device in the room, a deep problem is not being reached. The modality comparison sets the three side by side.
Many names are used interchangeably: SoftWave, StemWave, EPAT, BioFlex. Here is what each one actually is, taken from the manufacturers' own specifications and from FDA filings. Every figure can be checked at the source.
Focused and "focused-style" machines
| Machine | What it is | Maximum energy (mJ/mm²) | Maximum depth | FDA status |
|---|---|---|---|---|
| Storz DUOLITH SD1 | True focused shockwave, electromagnetic | Up to 0.55 (SEPIA handpiece) | Up to 125 mm | Premarket approval (PMA P080028) for chronic plantar fasciitis |
| SoftWave | Electrohydraulic; its FDA filings describe "unfocused pressure pulses" | 0.049 | 37 mm | Cleared for diabetic foot ulcers and burns; otherwise a Class I massager |
| StemWave | Electrohydraulic | Not published | Not published | Class I massager, no FDA clearance (per StemWave's own site) |
| EMS DolorClast Focused | Piezoelectric | 0.97 | 65 mm | Cleared for diabetic foot ulcers only |
| BTL Focused | Piezoelectric | Not published by BTL | About 35 mm | Class I massager |
| Richard Wolf PiezoWave2 | Piezoelectric | 0.40 in its FDA device listing (brochure claims 0.82) | 30 mm listed (100 mm claimed) | Class I massager listing |
Radial pressure wave machines, often sold as "shockwave"
| Machine | What it is | Maximum pressure | FDA status |
|---|---|---|---|
| Storz D-ACTOR 100 »ultra« (also sold as EPAT) | Radial pressure wave | 0.3 to 5 bar | Class I massager category, like every radial unit |
| Meditech BIOFLEX Shockwave | Radial pressure wave (its manual states "Shockwave type: Radial") | 5 bar | Class I massager listing |
| Chattanooga Intelect RPW 2 | Radial pressure wave | 5 bar | Class I massager listing |
| Zimmer enPuls | Radial pressure wave | "Equivalent to 5 bar" | Class I massager listing |
| EMS Swiss DolorClast Radial | Radial pressure wave | 4 bar | Class I massager; an older PMA for plantar fasciitis |
| Generic online "shockwave" units | Usually radial, unmeasured | Not measured | No FDA clearance |
Two points follow from these tables. First, every radial machine on the market, mine included, is regulated as a massager, and none of them reaches deep tissue the way a focused device does. Second, plantar fasciitis is the only musculoskeletal indication with FDA premarket approval for focused shockwave, and it belongs to the DUOLITH. Every other use, for every machine listed here, is off-label. That is legal and normal in clinical practice, but no clinic should imply FDA approval for treating a tendon, a hip or a back.
A note of fairness: a 2015 review found no evidence that one focused technology outperforms another, and concluded that dose and application matter more than the brand on the generator (Schmitz et al., Br Med Bull 2015). What a device can reach and how it is used are what separate them.
Whoever you see, these questions tell you quickly what you are being offered:
If those questions cannot be answered for your case, the treatment is a guess.
A situation I see every week: low back pain that travels down the leg. The GP orders an MRI, the right call, and the report reads "L4-L5 disc herniation." The diagnosis becomes sciatica.
That may be true. Nerve roots that contribute to the sciatic nerve exit at L4-L5, and a disc there can produce that pain. But was the patient examined? Was the pain reproduced, its pattern mapped, its aggravating and easing positions tested? That is how a diagnosis is actually reached, or at least narrowed to a short list.
Imaging shows what is there. It does not show what hurts. A Mayo Clinic-led review of 3,110 people with no back pain at all found disc bulges in 30% of twenty-year-olds and 84% of eighty-year-olds (Brinjikji et al., AJNR 2015). Much of what appears on an MRI is ordinary aging.
In my practice, a large share of patients who arrive with sciatic-type pain are not getting it from the spine. It comes from overloaded tissue around the hip: the piriformis, the gluteus medius and their neighbours, loaded by long hours of sitting or by running and training without the mobility work to match. The same principle applies throughout the body. I assess first, and I check whether the imaging matches what the examination finds. Often it does not, and the true source lies elsewhere.
Once the driver of the pain is clear, a course of care begins: most courses run 3 to 6 weekly sessions, and in my clinic the standard is six. Some patients get there sooner. It depends on the tissue, how long the problem has been present and how it responds.
Each visit is multimodal, about 30 minutes, delivered in a set order:
Paul Hobrough describes it as a dartboard: you work the outer rings, then you hit the bullseye.
This is where many clinics stop. The session ends and the patient leaves with a printout of clamshells and bridges. General rehabilitation has its place, but it is not built for that joint.
I use Functional Range Conditioning and am certified across the Functional Range Systems programs. We restore the joint's available motion, identify where the restrictions are, then rebuild it with loading chosen for that joint and that tissue. Shockwave, radial and EMTT help the tissue regenerate. The training raises its capacity so the problem does not return. That is the ADAPT Method in practice: LOAD > CAPACITY = PAIN, and the work is raising the capacity.
Good equipment in untrained hands is still a guess. Every device in my office is made by Storz Medical, a Swiss company that has built shockwave systems since 1987: the DUOLITH for focused shockwave, the D-ACTOR for radial pressure wave and the MAGNETOLITH for EMTT. I bought my equipment through CuraMedix, the US distributor of Storz Medical, and I attend their Institute for Tissue Regeneration and Repair conference in Boston, which I wrote about here.
My protocols were taught to me firsthand by my mentor, Paul Hobrough, one of the leading shockwave clinicians in the world, over the past two years (his interview with Storz Medical). In November 2026 I complete my certification with the American Shockwave Training Institute in Boulder. I am a member of the American Society for Medical Shockwave Treatment (ASMST).
Shockwave is one of the better-studied conservative treatments for chronic tendon, fascia and bone pain, but the evidence is uneven, and honest education means showing where it is strong and where it is not. In my practice it helps across most musculoskeletal conditions. How much it helps varies with the patient, the equipment, the clinician and the protocol, which is why the trials disagree as often as they do.
Every study behind these grades, positive and negative, is listed in the research library, and the evidence overview explains how each grade is set. For the questions patients ask before booking, see patient questions. Research supports shockwave for some of these conditions. It does not guarantee a result. Outcomes depend on an accurate diagnosis, how long the problem has been present and the work done between visits. The conditions I treat are listed at my practice.
Dr. Craig Patane, D.C., FRCms, is a sports chiropractor, running coach and running injury and movement specialist, and the creator of the ADAPT Method. He treats at Form & Function Chiropractic and Rehab in Brooklyn.
One study read properly, one protocol, and one note on running a cash practice. Every other Tuesday, free. Written by Dr. Craig Patane.
Shockwave sits inside a full rehab plan: see The ADAPT Method for clinicians
This site is for education, not medical advice. Shockwave, radial pressure wave, and EMTT are clinician-delivered therapies with specific indications and contraindications. Evidence strength varies by condition, and some uses are not FDA-cleared. Always consult a qualified healthcare provider about your individual situation.